Keysight Infiniivision 3000A X-Series Mixed Signal Oscilloscopes

Availability: In Stock / Short Lead Time
Only %1 left
£6,203.00

Keysight's InfiniiVision 3000A X-Series Mixed Signal Oscilloscopes redefine industry standards with a 6-in-1 instrument integration and exceptional performance. Offering fully upgradable bandwidth and additional digital channels, it guarantees maximum investment protection and delivers advanced capabilities within your budget.

Add to Wish List

InfiniiVision 3000A X-Series Mixed Signal Oscilloscopes

Keysight's InfiniiVision 3000A X-Series Mixed Signal Oscilloscopes redefine industry standards with a 6-in-1 instrument integration and exceptional performance. It boasts an industry-leading update rate of one million waveforms per second, 20 times faster than competitors, ensuring unmatched signal detail. Offering fully upgradable bandwidth and 16 additional digital channels, it guarantees maximum investment protection and delivers advanced capabilities within your budget.

Key Features

  • Bandwidth: 100 MHz ~ 1.5 GHz
  • Maximum Memory Depth: 4 Mpts
  • Maximum Sample Rate: 5 GSa/s
  • Number of Channels: 2 ~ 4 Analog plus 16 Digital
  • Waveform Update Rate: 1,000,000 Waveforms/sec
  • 6 Fully upgradeable instruments in 1 package
  • 20 MHz Dual-channel WaveGen with arbitrary waveform

For more information regarding pricing and lead times, get in touch at +44 (0) 800 583 4455 or sales@sjelectronics.co.uk.  

 

  3012A 3014A 3024A 3032A 3034A 3052A 3054A 3102A 3104A
Bandwidth1 (–3 dB) 100 MHz 200 MHz 350 MHz 500 MHz 1 GHz
Calculated Rise Time (10 to 90%) ≤ 3.5 ns ≤ 1.75 ns ≤ 1 ns ≤ 700 ps ≤ 450 ps
Input Channels 2 4 4 2 4 2 4 2 4
Maximum Sample Rate 4 GSa/s half channel, 2 GSa/s all channel  5 GSa/s half ch, 2.5 GSa/s all ch 
Maximum Memory Depth 2 Mpts half channels, 1 Mpts all channels (4 Mpts half channels, 2 Mpts all channels with DSOX3MEMUP upgrade option) 
Display Size and Type 8.5-inch WVGA display 
Waveform Update Rate > 1 million waveforms per second
Number of Active Probes Supported  In general, one for 2-channel models and two for 4-channel models. Contact us for specific configurations. 

 

Vertical System Analog Channels
Hardware Bandwidth Limits Approximately 20 MHz (selectable) 
Input Coupling AC, DC
Input Impedance Selectable: 1 MΩ ± 1% (14 pF), 50 Ω ± 1.5%
Input Sensitivity Range
100 MHz ~ 500 MHz models: 1 mV/div to 5 V/div2 (1 MΩ and 50 Ω)
1 GHz models: 1 mV/div to 5 V/div2 (1 MΩ), 1 mV/div to 1 V/div (50 Ω) 
Vertical Resolution 8 bits (measurement resolution is 12 bits with averaging)
Maximum Input Voltage 135 Vrms; 190 Vpk
Probing technology allows testing of higher voltages. For example, the included n2841A, N2842A or N2843A 10:1 probe supports testing up to 300Vrms 
Use this instrument only for measurements within its specified measurement category (not rated for CAT II, III, IV). No transient overvoltage allowed 
DC Vertical Gain Accuracy ± 2.0% full scale
DC vertical offset accuracy  ± 0.1 div ± 2 mV ± 1% of offset setting 
Channel-to-Channel Isolation  > 100:1 from DC to maximum specified bandwidth of each model (measured with same V/div and coupling on channels) 
Offset Range
± 2 V (1 mV/div to 200 mV/div)
± 50 V (> 200 mV/div to 5 V/div) 

 

Vertical System Digital Channels
Digital Input Channels 16 digital (D0 to D15. pod 1: D7 ~ D0, Pod 2: D15 ~ D8) 
Thresholds Threshold per pod
Threshold Selections TTL (+1.4 V), 5V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod) 
User-defined Threshold Range ± 8.0 V in 10 mV steps 
Maximum Input Voltage  ± 40 V peak CAT I
Threshold Accuracy ± (100 mV + 3% of threshold setting)
Maximum Input Dynamic Range ± 10 V about threshold 
Minimum Voltage Swing 500 mVpp
Input Impedance  100 kΩ ± 2% at probe tip
Input Capacitance ~8 pF
Vertical Resolution 1 bit

 

Horizontal System Analog Channels
  3012A 3014A 3024A 3032A 3034A 3052A 3054A 3102A 3104A
Time Base Range 5 ns/div to  50 s/div  2 ns/div to 50 s/div  1 ns/div to 50 s/div  500 ps/div to  50 s/div 
Time Base Accuracy 25 ppm ± 5 ppm per year (aging)
Time Base Delay Time Range Pre-trigger: Greater of 1 screen width or 250 μsPost-trigger: 1 to 500 s
Channel-to-Channel Deskew Range ± 100 ns
∆ Time Accuracy (using Cursors) ± (time base acc. x reading) ± (0.0016 x screen width) ± 100 ps 
Modes Main, zoom, roll, XY 
XY
On channels 1 and 2 only. Z Blanking on ext trigger input, 1.4 V threshold.
Bandwidth: Maximum bandwidth. Phase error at 1 MHz: < 0.5 degree. 

 

Horizon System Digital Channels
Minimum Detectable Pulse Width 5 ns
Channel-to-Channel Skew 2 ns (typical); 3 ns (maximum)

 

Acquisition System
  3012A 3014A 3024A 3032A 3034A 3052A 3054A 3102A 3104A
Maximum Analog Channels Sample Rate 4 GSa/s half channel interleaved, 2 GSa/s all channel  5 GSa/s half channel interleaved, 2.5 GSa/s all channel 
Maximum Analog Channels Record length  2 Mpts half channel interleaved, 1 Mpts all channel (standard)
4 Mpts half channel interleaved, 2 Mpts all channel (optional with DSOX3MEMUP  (-040)
Maximum Duration of Time Captured at Highest Sampling Rate  (All Analog Channels)  500 µs with 4M memory upgrade 400 µs with 4M memory upgrade
Maximum Digital Channels Sample Rate  1 GSa/s 1.25 GSa/s
Maximum Digital Channels Record Length 1 Mpts (standard - with digital channels only)
2 Mpts (optional with DSOX3MEMUP - with digital channels only) 
Modes
Normal: Default Mode
Peak Detect: Capture glitches as narrow as 250 ps at all time base settings 
Averaging: Selectable from 2, 4, 8, 16, 64, ... to 65,536 
High Resolution: Real-time boxcar averaging reduces random noise and effectively increases vertical resolution 12 bits of resolution when ≥ 10 μs/div at 4 GSa/s (5 GSa/s for 1 GHz models) or ≥ 20-μs/div at 2 GSa/s (2.5 GSa/s for 1 GHz models)
Segmented: Segmented memory optimizes available memory for data streams that have long dead times between activity. Maximum segments = 1000. Re-arm time = 1 μs  (minimum time between trigger events) 

 

Trigger System
Trigger Sources Analog channel (1~4), digital channel (D0~D15), line, external, WaveGen  (1 or mod) (FM/FSK)
Trigger Modes Normal (triggered): Requires trigger event for scope to trigger 
Auto: Triggers automatically in absence of trigger event 
Single: Triggers only once on a trigger event, press [Single] again for scope to find another trigger event, or press [Run] to trigger continuously in either Auto or Normal mode 
Force: Front panel button that forces a trigger 
Trigger Coupling
DC: DC coupled trigger
AC: AC coupled trigger, cutoff frequency: < 10 Hz  (internal); < 50 Hz (external) 
HF Reject: High frequency reject, cutoff frequency ~ 50 kHz 
LF Reject: Low frequency reject, cutoff frequency ~ 50 kHz 
Noise Reject: Selectable OFF or ON, decreases sensitivity 2x 
Trigger Holdoff Range 40 ns to 10.00 s

 

Trigger Sensitivity
Internal < 10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
External
200 mVpp from DC to 100 MHz
350 mVpp 100 MHz to 200 MHz 

 

Trigger Level Range
Any Channel ± 6 div from center screen
External
± 8 V 

 

Display Characteristics
Display 8.5-inch WVGA
Resolution 800 (H) x 480 (V) pixel format (screen area)
Graticules 8 vertical divisions by 10 horizontal divisions with intensity controls
Format YT, XY, and Roll
Maximum Waveform Update Rate > 1,000,000 waveforms/sec
Persistence Off, infinite, variable persistence (100 ms to 60 s)
Intensity Gradation  64 intensity levels