BINDER M 56 Drying and Heating Chamber

Availability: In Stock / Short Lead Time
  • Temperature range of +22°C above ambient temperature to a maximum of +300°C
  • APT.line™ preheating chamber technology
  • Forced convection

Binder M 56 Drying and Heating Chamber with Forced Convection and Advanced Program Functions

Advanced drying and heating chambers equipped with forced convection and sophisticated programmable features.

The BINDER M 56 material test chamber, capable of reaching up to +300°C and offering a range of programming options, is perfect for conducting material and aging tests, and its efficient high airflow system guarantees rapid heating.

Unit Dimensions

  • Housing dimensions (excl. fittings and connections): 560 mm (w) x 622mm (H) x 596mm (D)
  • Internal dimensions: 400 mm (w) x 440mm (H) x 335mm (D) 

Key Features

  • Temperature range of +22°C above ambient temperature to a maximum of +300°C
  • APT.line™ technology with increased heating output provides precision measurements when operating at high temperatures
  • Durable fan and adjustable speed 
  • Electromechanically controllable exhaust air flap
  • Easy-to-use touchscreen interface allows for real-time programming
  • 2 chrome-plated racks
  • Class 2 independent adjustable temperature safety device (DIN 12880) and class 3.1 independent temperature safety device (DIN 12880) with visual/ acoustic alarm available for selection on the controller
  • Backup sensor Pt 100 provides system failsafe function
  • Ethernet computer connectivity
  • Internal data logger. measured values can be viewed in open format from USB

 

Data
Designation  M056-230V
Article Number  9010-0371 
Option Model  Standard
Performance Data Temperature
Temperature range  +22°C above ambient temperature to 300°C
Temperature uniformity at 150°C  1.4 ± K
Temperature fluctuation at 150°C   0.2 ± K
Heating up time to 150°C  10 min
Recovery time after door was opened for 30 s at 150°C   4 min
Air Change Data
Air change rate at 100°C  170 x/h
Electrical Data
Rated Voltage  230 V
Power frequency  50/60 Hz
Nominal power   2 kW 
Unit fuse  10 A
Phase (Nominal voltage)  1~
Measures 
Interior volume  59 L
Net weight of the unit  43 kg
Load per rack  30 kg
Permitted load  70 kg 
Wall clearance back  160 mm
Wall clearance sidewise  100 mm
Housing Dimensions (Not including fittings and connections)
Width net  560 mm
Height net  622 mm
Depth net  596 mm 
Internal Dimensions
Interior width  400 mm
Interior height  440 mm
Interior depth  335 mm
Unit doors  1
Environment-Specific Data
Sound-pressure level  50 dB(A) 
Energy consumption at 150°C  320 Wh/h 
Fixtures
Number of shelves (std./max)  2/4