Fluke Calibration 1504 1 Channel Thermometer Readout

Availability: In Stock / Short Lead Time
£2,217.00
If you need more accuracy in a limited temperature range, the Model 1504 Tweener gives it to you as a thermistor readout.
Add to Wish List

1 Channel Thermometer for Thermistors

Best performance thermometers in their price range

  • Two Tweeners to choose from—reading PRTs or thermistors
  • Battery packs available
  • Best price/performance package

One of Hart’s best-selling products is the Tweener thermometer, and there’s a reason. No other company, not one, has a reference thermometer that comes close to the performance and features of the Tweener for anywhere near its price.

1504 Tweener Thermistor Readout

If you need more accuracy in a limited temperature range, the Model 1504 Tweener gives it to you as a thermistor readout. Thermistors are less fragile than PRTs and less likely to be impacted by mechanical shock. Thermistors are more sensitive to temperature, have faster response times, and come in many shapes for different applications.

Typical accuracy of 1504 is ±0.002 °C with a resolution of 0.0001 °C.

Software 

With our 9934 LogWare, both Tweener models may be used for real-time data acquisition. Collect data and analyze it graphically or statistically. Additionally, Tweeners may be used as reference thermometers with our MET/TEMP II software.

Battery pack

Want to take your Tweener on the road? Order Hart’s 9320A Battery Pack. Our battery gives you 36 hours between charges, allowing you to take your work wherever you need.

Calibration choices

Each Tweener and its accompanying probe (sold separately) have their own individual calibration reports. Overall system error can be calculated from the individual errors, rendering the added cost of system data unnecessary. However, for those requiring it, system data is available at two or more temperatures of your choice. (See Calibration Model 1929-X )


Specifications

1502A 1504
Temperature Range −200 °C to 962 °C (−328 °F to 1764 °F) Any thermistor range
Resistance Range 0W to 400W, auto-ranging 0W to 1 MW, auto-ranging
Probe Nominal RTPW: 10W to 100W
RTD, PRT, or SPRT
Thermistors
Characterizations ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11
IPTS-68: R0, a, d, a4, and c4
Callendar-Van Dusen: R0, a, d, and b
Steinhart-Hart thermistor polynomial
Callendar-Van Dusen: R0, a, d, and b
Resistance Accuracy
(ppm of reading)
0W to 20W: 0.0005W
20W to 400W: 25 ppm
0W to 5 KW: 0.5W
5 KW to 200 KW: 100 ppm
200 KW to 1 MW: 300 ppm
Temperature Accuracy ±0.004 °C at −100 °C
±0.006 °C at 0 °C
±0.009 °C at 100 °C
±0.012 °C at 200 °C
±0.018 °C at 400 °C
±0.024 °C at 600 °C
±0.002 °C at 0 °C
±0.002 °C at 25 °C
±0.004 °C at 50 °C
±0.010 °C at 75 °C
±0.020 °C at 100 °C
(Using 10 KW thermistor sensor, a=0.04. Does not include probe uncertainty or characterization errors.)
Operating Temperature Range 16 °C to 30 °C 13 °C to 33 °C
Resistance Resolution 0W to 20W: 0.0001W
20W to 400W: 0.001W
0W to 10 KW: 0.01W
10 KW to 100 KW: 0.1W
100 KW to 1 MW: 1W
Temperature Resolution 0.001 °C 0.0001 °C
Excitation Current 0.5 and 1 mA, user selectable, 2 Hz 2 and 10 mA, automatically selected
Measurement Period 1 second
Digital Filter Exponential, 0 to 60 seconds time constant (user selectable)
Probe Connection 4-wire with shield, 5-pin DIN connector
Communications RS-232 serial standard
IEEE-488 (GPIB) optional
Display 8-digit, 7-segment, yellow-green LED; 0.5-inch-high characters
Power 115 VAC (±10 %), 50/60 Hz, 1 A, nominal
230 VAC (±10 %), 50/60 Hz, 1 A, nominal, specify
Size (HxWxD) 61 x 143 x 181mm (2.4 x 5.6 x 7.1 in)
Weight 1.0 kg (2.2 lb.)
Probes from Hart 5615, 5627, 5626, 5628, 5622 5640-44, 5610-65
Calibration ISO 17025-accredited calibration provided
Temperature ranges and accuracy may be limited by the sensor you use.