Fluke Calibration 9102S-256 Handheld Dry-Well

Availability: In Stock / Short Lead Time
£3,064.00
Handheld Dry-Well, 220 V / 50/60 Hz
Add to Wish List

For work in the temperature range of –10°C to 122°C, our Model 9102S dry-well is another first in the industry, featuring stability to ±0.05°C.

This dry-well is only four inches high and six inches wide, achieves temperatures as low as –10°C, includes a NIST-traceable calibration, and is stable to ±0.05°C. The Model 9102S is excellent for dial gauges, digital thermometers, bulb switches, and other sensors that need calibration below ambient.

The 9102S has two wells so you can use one for a reference thermometer to increase accuracy. Both wells are 1/2 inch in diameter, and each has inserts available for almost any sensor size. The 9102S also has a battery pack option that gives you approximately four hours of field use when AC power is unavailable.



Fluke Calibration 9100 Handheld Dry-Well - Series

World’s smallest, lightest and most portable dry-wells

  • Smallest dry-wells in the world
  • Ranges from –10°C to 375°C
  • Accuracy to ±0.25°C, stability of ±0.05°C at 0°C
  • RS-232 interface with 9930 Interface-it v3.81 software

Fluke Calibration's line of portable dry-wells is incredible. They’re the smallest, lightest, and most portable dry-wells in the world. And now they're better than ever!


Specifications 9100S 9102S
Range 35°C to 375°C (95°F to 707°F) –10°C to 122°C (14°F to 252°F) at 23°C ambient
Accuracy ±0.25°C at 50°C; ±0.25°C at 100°C; ±0.5°C at 375°C ±0.25°C
Stability ±0.07°C at 50°C; ±0.1°C at 100°C; ±0.3°C at 375°C ±0.05°C
Well-to-well uniformity ±0.2°C with sensors of similar size at equal depths within wells
Heating times 35°C to 375°C: 9.5 minutes Ambient to 100°C: 10 minutes
Stabilization 5 minutes 7 minutes
Cooling times 375°C to 100°C: 14 minutes Ambient to 0°C: 10 minutes
Well depth 102 mm (4 in); 1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep 102 mm (4 in)
Removable inserts N/A Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included]
Power 115 V AC (±10%), 55–65 Hz, 1.5 A or 230 V AC (±10%), 0.8 A, 45–55 Hz, 175 W 94-234 V AC (±10%), 50/60 Hz, 60 W; or 12 V DC
Size (H x W x D) 57 x 125 x 150 mm (2.25 x 4.9 x 6.1 in) 99 x 140 x 175 mm (3.9 x 5.5 x 6.9 in)
Weight 1 kg (2 lb 3 oz) 1.8 kg (4 lb)
Computer interface RS-232 included with free Interface-it software
NIST-traceable calibration Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C and 375°C Data at –10°C, 0°C, 25 °C, 50°C, 75 °C,100°C and 122°C